KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern
10 KEN-TM / KET_TM-BA-e-1420 1.2.2 KEN-TM 1. Protective conductor 2. Electric Supply 3. Ethernet 4. 3IN/OUT 5. RS232 6. USB port 7.
100 KEN-TM / KET_TM-BA-e-1420 The setting options for the application “Parts counting“, are with a few exceptions identical to those of the appl
KEN-TM / KET_TM-BA-e-1420 101 Counting with reference optimisation For setting reference single weight (see chap. 10.3.1or chap. 10.3.2), co
102 KEN-TM / KET_TM-BA-e-1420 10.2.2 Minimum reference single weight Tap <Minimum reference mass>. Select desired setting. Enteri
KEN-TM / KET_TM-BA-e-1420 103 10.3 Carry out parts counting 10.3.1 Determination of the reference piece weight by weighing Determining referenc
104 KEN-TM / KET_TM-BA-e-1420 The weighing scale determines the reference single weight and displays the result on the info field. Remove
KEN-TM / KET_TM-BA-e-1420 105 10.3.2 Enter reference single weight as numeric value Set reference: Press button. Enter known referen
106 KEN-TM / KET_TM-BA-e-1420 10.4.1 Load reference single weight from database. Load reference If a reference single weight was previo
KEN-TM / KET_TM-BA-e-1420 107 10.4.2 Allocating reference single weight in database to a product Procedure 1 via function keys: Press or fu
108 KEN-TM / KET_TM-BA-e-1420 Procedure 2 via menu: For determining a reference single weight see chap. 10.3.1 or chap.10.3.2 Press to i
KEN-TM / KET_TM-BA-e-1420 109 11 Percent determination Percent weighing allows to display weight in percent, in relation to a reference weight (
KEN-TM / KET_TM-BA-e-1420 11 Pin allocation: RS 232 / BUS Module (option) Pin1 – B Pin2 – RxD Pin3 – TxD Pin4 – A Pin5 – GND Pin6 – +5VDC Addit
110 KEN-TM / KET_TM-BA-e-1420 11.2 Application specific settings Tap icon in the bottom left corner of the window or call it up from the m
KEN-TM / KET_TM-BA-e-1420 111 11.3 How to carry out percentage calculation 11.3.1 Determination of the reference weight by weighing Determining
112 KEN-TM / KET_TM-BA-e-1420 Percent weighing Place goods to be weighed on balance. The weight of the item to be weighed is displayed in p
KEN-TM / KET_TM-BA-e-1420 113 The weight will be imported as reference (100%) and displayed in the info field. Percent weighing Place goods
114 KEN-TM / KET_TM-BA-e-1420 12 Formulation This function is used to weigh several related components one after another. The weighing result
KEN-TM / KET_TM-BA-e-1420 115 The function keys below as well as a special info field for Recipetion are enabled by default. For application sp
116 KEN-TM / KET_TM-BA-e-1420 12.2 Application specific settings Tap icon in the bottom left corner of the window or call it up from the me
KEN-TM / KET_TM-BA-e-1420 117 12.3 Defining recipe in database Call up recipe database: Call up menu by Tap <Databases> T
118 KEN-TM / KET_TM-BA-e-1420 The menu will appear. Tap the individual icons to call up and edit the individual settings as described below.
KEN-TM / KET_TM-BA-e-1420 119 3. Component This is used to define the individual components of the recipe. Go to menu and tap < Component
12 KEN-TM / KET_TM-BA-e-1420 2 Control elements and program structure 2.1 Keyboard overview Fig.: KET-TM Fig.: KEN-TM
120 KEN-TM / KET_TM-BA-e-1420 Go to menu and tap <Products> . Select component from database “Product“ Go to menu and tap <Mass
KEN-TM / KET_TM-BA-e-1420 121 Define parameters for additional components: After defining all parameters for the first component return to
122 KEN-TM / KET_TM-BA-e-1420 The number of components and the total weight will be displayed. 4. Type of charge Go to the menu
KEN-TM / KET_TM-BA-e-1420 123 By ingredient Based on the recipe, several batches will be produced that contain only one of the components of th
124 KEN-TM / KET_TM-BA-e-1420 12.4 Process recipe from database Condition: The user must be logged on at least at <user> level. Re
KEN-TM / KET_TM-BA-e-1420 125 Press function key and select desired recipe such as MiHo-Creme (see chap. 12.3) from database. Place
126 KEN-TM / KET_TM-BA-e-1420 The bargraph of the weighing aid runs from the left to the right and progresses according to the weight placed on
KEN-TM / KET_TM-BA-e-1420 127 After weighing all components you will see the message <End of process >. Confirm message by . Thi
128 KEN-TM / KET_TM-BA-e-1420 12.5 Log recipe When < > is enabled (see chap.12.2) a log will be issued automatically after every formula
KEN-TM / KET_TM-BA-e-1420 129 To <Open / Print> keep your finger pressed on the desired report until the context menu is displayed. Th
KEN-TM / KET_TM-BA-e-1420 13 KET-TM KEN-TM 1 - Touch-free sensors. Adjustable function, see chap. 7.4.2. Operate this function by moving yo
130 KEN-TM / KET_TM-BA-e-1420 Printout example default log (KERN YKB-01N):
KEN-TM / KET_TM-BA-e-1420 131 13 Animal weighing This function is used to weigh unstable articles such as live animals. There are two different
132 KEN-TM / KET_TM-BA-e-1420 13.2 Application specific settings Tap icon in the bottom left corner of the window or call it up from the m
KEN-TM / KET_TM-BA-e-1420 133 13.2.1 Average time This function is used to set the duration for the weighing cycle based on which average is calc
134 KEN-TM / KET_TM-BA-e-1420 13.3 Dynamic weighing using manual start Condition: For auto start disabled see chap. 13.2.2. When using a
KEN-TM / KET_TM-BA-e-1420 135 13.4 Dynamic weighing with automatic start System requirements: Auto start enabled, see chap. 13.2.2. For LO thre
136 KEN-TM / KET_TM-BA-e-1420 14 Database Call up database menu: Press . Tap <Datenbases > button. The menu for databas
KEN-TM / KET_TM-BA-e-1420 137 14.1 Configuring database (administrator only) Go to menu (See chap. 14) and tap < Database configurati
138 KEN-TM / KET_TM-BA-e-1420 14.1.1 Accessibility of databases This is used to determine which menu items are to be available in the <Data
KEN-TM / KET_TM-BA-e-1420 139 Tap <Name>. Enter name (such as cat. 1) in editing window and import by tapping . 1. Enable displa
14 KEN-TM / KET_TM-BA-e-1420 2.2 Overview of display The coloured touch screen is a touch-sensitive display. The touch screen does not only dis
140 KEN-TM / KET_TM-BA-e-1420 14.1.3 Products / Operators / Clients This is used to determine which data is to be made available in the databa
KEN-TM / KET_TM-BA-e-1420 141 14.1.4 Extra variables This is used to define extra variables that are not available on the list of variables, see
142 KEN-TM / KET_TM-BA-e-1420 Display example on calling up <Database> / <Product>: 14.1.6 Importing / exporting databas
KEN-TM / KET_TM-BA-e-1420 143 14.2 Search for Data Record / Add Data Record Call up < Databases >, see chap. 14. Select
144 KEN-TM / KET_TM-BA-e-1420 2. Search data record by code Tap In the displayed editing window, enter <Code> and confirm b
KEN-TM / KET_TM-BA-e-1420 145 14.4 Deleting data record (administrator only) Call up < Databases >, see chap. 14.
146 KEN-TM / KET_TM-BA-e-1420 14.5 Print data record Call up < Databases >, see chap. 14. Select <Product&
KEN-TM / KET_TM-BA-e-1420 147 14.6 Basic data record operations via context menu 1. Context menu “databases“ Call up < Databases &g
148 KEN-TM / KET_TM-BA-e-1420 2. Context menu “data record“ Call up < Databases >, see chap. 14. Select desired submenu
KEN-TM / KET_TM-BA-e-1420 149 14.7 Edit databases (administrator only) Sequence of operations: Open database menu < Database>, see chap
KEN-TM / KET_TM-BA-e-1420 15 2. Range 5 Current measuring value. The non-calibrated part of weighing scales with type approval is highlighted. 6
150 KEN-TM / KET_TM-BA-e-1420 14.7.2 Product database Parameter overview: Name Name of product Description Additional description of produ
KEN-TM / KET_TM-BA-e-1420 151 Ingredients Editing field for entering recipe component Label Assign single label sample to product Label K Assi
152 KEN-TM / KET_TM-BA-e-1420 14.7.3 Client database Parameter overview: Name Name of client Code Client code Tax ref. number Tax ID numb
KEN-TM / KET_TM-BA-e-1420 153 14.7.5 Database “Recipes“ Parameter overview: Name Name of recipe Code Code of recipe Ingredients Define comp
154 KEN-TM / KET_TM-BA-e-1420 14.7.8 Database “Storage“ Parameter overview: Name Name of storage space Code Code of storage space Descrip
KEN-TM / KET_TM-BA-e-1420 155 15 Logs How to call up log menu: Press . Tap <Reports> button. The menu for logs will be
156 KEN-TM / KET_TM-BA-e-1420 15.1 Enabling accessibility to logs This is used to determine which menu items are to be accessible in menu <
KEN-TM / KET_TM-BA-e-1420 157 15.2 Deleting older data (advanced users only) Tap < Delete older data >. Enter date up to wh
158 KEN-TM / KET_TM-BA-e-1420 15.3 Weighing logs Tap <Reports>, see chap.15. Select <Reports from weighing re
KEN-TM / KET_TM-BA-e-1420 159 15.3.1 Alibi memory If required by statutory provisions, all weighing activities may be traced in the alibi memory
16 KEN-TM / KET_TM-BA-e-1420 2.3 Program structure 2.3.1 Menu overview The menu is split into the following blocks: WEIGHING see chap. 7.1
160 KEN-TM / KET_TM-BA-e-1420 15.3.2 Filter This function is used to filter the printouts of alibi log files by specific criteria. Selec
KEN-TM / KET_TM-BA-e-1420 161 Tap Criteria such as starting date. Filter enabled Filter disabled Tap starting date. Enter year, month
162 KEN-TM / KET_TM-BA-e-1420 Settings default log: ---------------------------------------- Log of weighing activities ----------------------
KEN-TM / KET_TM-BA-e-1420 163 15.3.4 Display weighing procedures as a diagram Select <Weigments chart >, see chap. 15.3.
164 KEN-TM / KET_TM-BA-e-1420 15.3.5 Export weighing data to USB storage medium For selecting <Export the weighing database to a file
KEN-TM / KET_TM-BA-e-1420 165 15.3.6 Counter for weighing activities This function is used to edit the counter for weighing activities. Fo
166 KEN-TM / KET_TM-BA-e-1420 16 Communication 16.1 Overview interface commands The weighing scale recognises the following commands. Comman
KEN-TM / KET_TM-BA-e-1420 167 16.2 General response format Commands Function XX˽A CR LF Command accepted; executing command XX˽D CR LF Execut
168 KEN-TM / KET_TM-BA-e-1420 16.3 Detailed information about interface logs 16.3.1 Zeroing Befehl: Z CR LF Possible responses: Z˽A CR LF
KEN-TM / KET_TM-BA-e-1420 169 16.3.3 Retrieve tare value Command: OT CR LF Answer: 1 2 3 4 - 12 13 14 - 16 17 18 19 O T ˽ Tare ˽ Un
KEN-TM / KET_TM-BA-e-1420 17 2.3.2 Navigation in the menu Call up menu Scroll forward in menu Scroll backwards in menu Scroll „down“. Conf
170 KEN-TM / KET_TM-BA-e-1420 Example: S CR LF Sent command: Send stable weighing value in standard weighing unit S ˽ A CR LF Command accepted;
KEN-TM / KET_TM-BA-e-1420 171 16.3.7 Send weighing values for all platforms in default weighing unit immediately Command: SIA CR LF Possible re
172 KEN-TM / KET_TM-BA-e-1420 16.3.9 Send weighing value immediately in current weighing unit Command: SUI CR LF Possible responses: SUI˽I
KEN-TM / KET_TM-BA-e-1420 173 16.3.12 Start continuous output in current weighing unit Command: CU1 CR LF Possible responses: CU1˽I CR LF Com
174 KEN-TM / KET_TM-BA-e-1420 16.3.16 Retrieve value for “Lower Threshold“ Command: ODH CR LF Possible responses: DH˽MASA CR LF Command
KEN-TM / KET_TM-BA-e-1420 175 16.4 Data format 1 2 3 4-12 13 14-16 17 18 stability ˽ Signs Measuring Value ˽ Unit CR LF Stability sy
176 KEN-TM / KET_TM-BA-e-1420 17 Servicing, maintenance, disposal Before any maintenance, cleaning and repair work disconnect the appliance fro
KEN-TM / KET_TM-BA-e-1420 177 18 Troubleshooting guide / error messages Possible causes of errors: In case of an error in the program process, bri
178 KEN-TM / KET_TM-BA-e-1420 19 Appendix A – Variables for printouts 19.1 List of variables Symbol Description {0} 1) Default printout in a
KEN-TM / KET_TM-BA-e-1420 179 {31} Number of platform {32} Manufacturer’s number {33} Weighing scale division {34} Range {35} Parts counting:
18 KEN-TM / KET_TM-BA-e-1420 Exit menu / back to weighing mode. The display returns one step to previous view by tapping Press Back to Main
180 KEN-TM / KET_TM-BA-e-1420 {77} User: Authorisations {80} Packaging: Name {81} Packaging: Code {82} Packaging: Weight {85} Customer: Nam
KEN-TM / KET_TM-BA-e-1420 181 {137} Target storage place: Description {140} Net weight in adjusting unit: Sum {141} Ancillary display: WD {142}
182 KEN-TM / KET_TM-BA-e-1420 {187} Measurements: Difference {190} 3) Comparator: Log number {191} 3) Comparator: Starting date {192} 3) Com
KEN-TM / KET_TM-BA-e-1420 183 {234} Recipe: Status {235} Recipe: Min {236} Recipe: Max {237} Recipe: Code of ingredient {240} Recipe log: St
184 KEN-TM / KET_TM-BA-e-1420 19.2 Format variables The user has the option to format randomly any number variables and text variables that ca
KEN-TM / KET_TM-BA-e-1420 185 Special characters for formatting: Character Description Example , Delimiter for variables with left-justified al
186 KEN-TM / KET_TM-BA-e-1420 Examples: CODE Description {7:V6.3} Net weight as EAN 13 (code of 6 characters) {7:V7.3} Net weight as EAN 13
KEN-TM / KET_TM-BA-e-1420 187 19.2 Mathematic formula The function for mathematic formulas with variable <{5} Mathematic Formula> makes it
188 KEN-TM / KET_TM-BA-e-1420 There are additional functions that modify text values. These functions allow you to amend text during its issue:
KEN-TM / KET_TM-BA-e-1420 189 20 Appendix B – list “Function Keys“ Symbol Description ENTER Zeroing Taring Set tare Parameters Application
KEN-TM / KET_TM-BA-e-1420 19 2.4 Authorisation levels The software is designed for users with different access rights: Administrator and various
190 KEN-TM / KET_TM-BA-e-1420 Select product by code Select packaging Select packaging by name Select packaging by code Select customer Se
KEN-TM / KET_TM-BA-e-1420 191 Permit pouring Disable tare Retrieve tare Change unit Change platform Edit universal variable 1 Edit univer
192 KEN-TM / KET_TM-BA-e-1420 Density determination with pycnometer Density determination for porous solids Edit quantity of labels Edit qu
KEN-TM / KET_TM-BA-e-1420 193 21 Installation display unit / platform Installation / configuration of a weighing system must be carried out by a
194 KEN-TM / KET_TM-BA-e-1420 21.3 How to connect the platform Disconnect the display unit from the power supply. Open housing To shiel
KEN-TM / KET_TM-BA-e-1420 195 With 4 conductors: Printed Circuit Board Load cell Remarks E SHIELD REF+ - JP1 jumpered REF- - JP2 jumpe
196 KEN-TM / KET_TM-BA-e-1420 21.4 Configure display unit Before configuring a display unit make sure you know the required configuration data,
KEN-TM / KET_TM-BA-e-1420 197 General parameters Tap < Global >. The menu will be displayed; see table below. 10. To call up or chang
198 KEN-TM / KET_TM-BA-e-1420 Tab. 10: Number of platforms Enter number of platforms Weighing module type Standard Weighing scale with NT com
KEN-TM / KET_TM-BA-e-1420 199 Manufacturer parameters 1. Tap < Factory parameters>. For configuring one platform, the submenu will be dis
2 KEN-TM / KET_TM-BA-e-1420 GB KERN KEN-TM / KET-TM Version 2.0 11/2014 Operating and installation instructions Display unit Contents 1
20 KEN-TM / KET_TM-BA-e-1420 3 Basic instructions 3.1 Proper use The display unit acquired by you is used in combination with a load cell and
200 KEN-TM / KET_TM-BA-e-1420 21.4.1 Entering metrology / configuration data Tap < Metrology > and the list of parameters for con
KEN-TM / KET_TM-BA-e-1420 201 List of parameters “Configuration“ (Display example single range scale 6 kg): Parameters Default setting Adjustable
202 KEN-TM / KET_TM-BA-e-1420 21.4.2 Adjusting a weighing system Tap < Adjustment > The menu will be displayed Determine initial m
KEN-TM / KET_TM-BA-e-1420 203 Tap < Adjustment > Ensure that there are no objects on the weighing plate. Acknowledge with .
204 KEN-TM / KET_TM-BA-e-1420 21.4.3 How to linearize a weighing system Linearity shows the greatest deviation of a weight display on the scal
KEN-TM / KET_TM-BA-e-1420 205 Tap <Estimate>. Display example max 6 kg: 1. Enter weight value for first adjustment point (1/6 max) an
206 KEN-TM / KET_TM-BA-e-1420 Correct linearization points: Tap < Corrections > The linearization points including related deviati
KEN-TM / KET_TM-BA-e-1420 207 21.4.4 Display configuration data Tap < Information >. The configuration data will be displayed 21
208 KEN-TM / KET_TM-BA-e-1420 21.4.6 Geo value The geographical value is used to adapt the weighing scale to local gravity conditions. Set co
KEN-TM / KET_TM-BA-e-1420 209 Set geo value for adjustment location and installation location. Tap < Global >. The menu will be disp
KEN-TM / KET_TM-BA-e-1420 21 3.2 Improper Use Do not use weighing system for dynamic weighing. In the event that small quantities are removed or
210 KEN-TM / KET_TM-BA-e-1420 1. Gravitational acceleration method Enter geo value for adjustment location Enter geo value for insta
KEN-TM / KET_TM-BA-e-1420 211 2. Latitude method Enter gravitational acceleration for adjustment location. Enter latitude for insta
212 KEN-TM / KET_TM-BA-e-1420 22 Conformity explanation/ test certificate KERN & Sohn GmbH D-72322 Balingen-Frommern Postbox 4052 E-Mail: i
22 KEN-TM / KET_TM-BA-e-1420 4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this
KEN-TM / KET_TM-BA-e-1420 23 6 Unpacking, Setup and Commissioning 6.1 Installation Site, Location of Use The display units are designed in a way
24 KEN-TM / KET_TM-BA-e-1420 6.4 How to connect the platform See chap. 21 “Installing display unit / platform“. Please also follow the instruc
KEN-TM / KET_TM-BA-e-1420 25 + Log in , Tap <Log in> button. <Operators> window appears. The weighing scale is supplied withou
26 KEN-TM / KET_TM-BA-e-1420 + Setting user language On delivery the display is set to German. For setting additional languages see chap.7.8.1.
KEN-TM / KET_TM-BA-e-1420 27 6.9 Basic Operation + Stabilisation requires a certain warm-up time. + For further information on specific settin
28 KEN-TM / KET_TM-BA-e-1420 6.9.3 Selecting weighing unit Tap unit to select from the displayed menu. Not all weighing units are avail
KEN-TM / KET_TM-BA-e-1420 29 6.9.4 Weighing with tare Taring The dead weight of any weighing container may be tared away by pressing a button,
KEN-TM / KET_TM-BA-e-1420 3 6.9.1 Simple weighing ...
30 KEN-TM / KET_TM-BA-e-1420 Delete tare 1. Unload weighing plate and press . The (Net) indicator turns off, the zero display shows. 2. W
KEN-TM / KET_TM-BA-e-1420 31 7 System settings (menu) The system settings are used to customise the behaviour of the weighing scales to your requ
32 KEN-TM / KET_TM-BA-e-1420 7.1 Weighing This function is used to customise the behaviour of the weighing scale to the ambient conditions of
KEN-TM / KET_TM-BA-e-1420 33 7.1.1 Median filter The median filter removes shocks such as mechanical vibrations. Tap <Medianfilter> The
34 KEN-TM / KET_TM-BA-e-1420 7.1.3 Auto Zero The automatic zero point adapter (auto zero) provides continuous correction for minor weight var
KEN-TM / KET_TM-BA-e-1420 35 7.2 Communication parameters Via the interfaces weighing data may be exchanged with connected peripheral devices. C
36 KEN-TM / KET_TM-BA-e-1420 7.2.2 Ethernet / IP Tap < Ethernet > The configuration options below will be displayed: DHCP yes no IP-a
KEN-TM / KET_TM-BA-e-1420 37 7.3 Devices Tap <Devices>. Select peripheral device to be configured.
38 KEN-TM / KET_TM-BA-e-1420 7.3.1 PC Select / configure computer interface Tap < Computer > Configuration options will be displayed
KEN-TM / KET_TM-BA-e-1420 39 3. Continuous transmission Setting options: no Continuous data transmission disabled yes Continuous da
4 KEN-TM / KET_TM-BA-e-1420 8 APPLICATION SETTINGS ... 79 8.1
40 KEN-TM / KET_TM-BA-e-1420 5. E2R system Tap < E2R System >. The settings <System enabled> / <Product selection blocked&
KEN-TM / KET_TM-BA-e-1420 41 7.3.2 Printer Selecting / configuring printer interface Tap < Printer > Configuration options will be displ
42 KEN-TM / KET_TM-BA-e-1420 3. Defining logs Display example: Default log “Weighing“ Tap <Printouts>. Select log type (such as w
KEN-TM / KET_TM-BA-e-1420 43 7.3.3 Barcode scanner Select / configure interface The barcode scanner allows fast access to: Product Custom
44 KEN-TM / KET_TM-BA-e-1420 2. Prefix / suffix Tap < Prefix / Suffix >. Enter values (hexadecimal) for prefix / suffix in input window
KEN-TM / KET_TM-BA-e-1420 45 4. Test Tap < Test >. Check error-free functioning of barcode scanner. 7.3.4 Transponder card reader Sele
46 KEN-TM / KET_TM-BA-e-1420 7.3.5 Additional display Tap < Additional display > 1. Port Tap <Port>. The available interfa
KEN-TM / KET_TM-BA-e-1420 47 7.4 Configure display This function allows the user to customise the terminal to his/her requirements. Tap
48 KEN-TM / KET_TM-BA-e-1420 Main display sample (middle) Select information texts in info field The info field is displayed in the applicatio
KEN-TM / KET_TM-BA-e-1420 49 Table 2: Default settings “information texts“ Weighing: {40:Product:,-15}{50} {40:Tare:,-15}{9}{11} {40:Gross:,-15}{
KEN-TM / KET_TM-BA-e-1420 5 13.4 Dynamic weighing with automatic start ...
50 KEN-TM / KET_TM-BA-e-1420 7.4.2 Selecting function keys Function keys allow direct access to frequently used functions and settings of the e
KEN-TM / KET_TM-BA-e-1420 51 Table 3: Default settings “function keys“ Weighing: F1-key Search product F2-key Select customer F3-key Set
52 KEN-TM / KET_TM-BA-e-1420 Screen key 2 Select dispensing Screen key 3 Start Screen key 4 Stop Screen key 5 Failure Formulation:
KEN-TM / KET_TM-BA-e-1420 53 7.4.3 Display all platforms When used as weighing system with several platforms, the display for each platform can
54 KEN-TM / KET_TM-BA-e-1420 7.4.4 Select bargraph type Tap < Bargraph >. The available bargraph types will be displayed: None – ba
KEN-TM / KET_TM-BA-e-1420 55 2. Fast weighing Function: Eight red and three green pilot lights indicate whether the load is within the toleranc
56 KEN-TM / KET_TM-BA-e-1420 Tab. 4: Operating mode for thresholds MIN, MAX Stable – pilot light for thresholds MIN, MAX will be visible aft
KEN-TM / KET_TM-BA-e-1420 57 Settings: Tap < Signalling checkweighing ranges >. Menu will appear, see table. 4. 4. Linearity Functio
58 KEN-TM / KET_TM-BA-e-1420 Settings: Tap < Linear >. Menu will appear, see table. 5. Tab. 5: Pilot light colour for MIN threshold
KEN-TM / KET_TM-BA-e-1420 59 7.5 I/O settings Tap < Inputs / Outputs >.
6 KEN-TM / KET_TM-BA-e-1420 16.3.7 Send weighing values for all platforms in default weighing unit immediately ... 171 16.3.8
60 KEN-TM / KET_TM-BA-e-1420 7.5.1 Inputs For pin allocation IN1 – IN4, see chap. 1.2 Tap < Inputs >. Select input to conf
KEN-TM / KET_TM-BA-e-1420 61 7.5.2 Outputs For pin allocation OUT1 – OUT4, see chap. 1.2 Tap < Outputs >. Select output to co
62 KEN-TM / KET_TM-BA-e-1420 7.6 Authorisations Function <Access Level> is only available when you are logged on as <Administrator&
KEN-TM / KET_TM-BA-e-1420 63 2. Date and time This function is used to grant the user access to <Date and Time>. Tap < Date and time
64 KEN-TM / KET_TM-BA-e-1420 4. Editing database This function is used to grant rights that are supposed to be available to the user for editi
KEN-TM / KET_TM-BA-e-1420 65 5. Select database This function is used to grant rights that are to be available to the user for viewing the respe
66 KEN-TM / KET_TM-BA-e-1420 7.7 Weighing Units This function is used to determine which weighing units are to be used for the operation of th
KEN-TM / KET_TM-BA-e-1420 67 2. Starting unit This function is used to set the unit to be displayed when the weighing scale is started. Tap <
68 KEN-TM / KET_TM-BA-e-1420 7.8 General parameters This is used to set parameters influencing the operation of the weighing scale, such as us
KEN-TM / KET_TM-BA-e-1420 69 7.8.2 Enter date / time, select display format Tap < Date and time >. The individual settings can be call
KEN-TM / KET_TM-BA-e-1420 7 1 Technical data KERN KET-TM KEN-TM Display (segments) 7-digit 7-digit Resolution (verified) 6000 e 6000 e Reso
70 KEN-TM / KET_TM-BA-e-1420 2. Selecting date format Tap < Date format > Import desired format by tapping. 3. Selecting time forma
KEN-TM / KET_TM-BA-e-1420 71 7.8.3 Signal tone when pressing button Every time a key is pressed, a brief audio signal will be sounded for confirm
72 KEN-TM / KET_TM-BA-e-1420 7.8.4 Set the touch screen alignment When the alignment of the touch-sensitive ranges of the display does not matc
KEN-TM / KET_TM-BA-e-1420 73 7.8.5 Setting brightness of display Tap < Screen brightness >. The currently set brightness will be d
74 KEN-TM / KET_TM-BA-e-1420 7.8.7 Setting “Log on required“ Tap <Login in required >. To force the user to log on after starting up
KEN-TM / KET_TM-BA-e-1420 75 2. Standard Settings Tap < Set Default >. To reset the default setting for the start logo, tap 7.8.9
76 KEN-TM / KET_TM-BA-e-1420 7.9 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each display
KEN-TM / KET_TM-BA-e-1420 77 Notes on verified weighing systems In calibrated weighing systems the adjustment switch used for adjustments is loc
78 KEN-TM / KET_TM-BA-e-1420 7.11 Weighing scale info This function is used to call up weighing scale information. Tap < Info >
KEN-TM / KET_TM-BA-e-1420 79 8 Application settings Overview of available applications: Weighing see chap. 9. Parts counting see chap. 10. P
8 KEN-TM / KET_TM-BA-e-1420 1.1 Dimensions KET-TM: KEN-TM:
80 KEN-TM / KET_TM-BA-e-1420 Which applications are to be available on the menu is defined under menu item < functions availability>.
KEN-TM / KET_TM-BA-e-1420 81 2. Call up application on menu For menu access press . Tap < Working modes >. Tap desire applica
82 KEN-TM / KET_TM-BA-e-1420 8.2 Application specific settings Application specific settings are available to customise the application accord
KEN-TM / KET_TM-BA-e-1420 83 1. Tap the icon on the bottom left corner of the window. This action will bring up the application specific s
84 KEN-TM / KET_TM-BA-e-1420 2. For menu access please press . Tap < Working modes >. Tap desired application (such as weigh
KEN-TM / KET_TM-BA-e-1420 85 9 Weighing The default setting for the weighing scale in the application is weighing. How to carry out a simple we
86 KEN-TM / KET_TM-BA-e-1420 9.1 Setting storage conditions for alibi memory This function is used to determine under which conditions weighi
KEN-TM / KET_TM-BA-e-1420 87 Under the menu item <Reports> <Reports from weighing records> <Weighings / Alibi
88 KEN-TM / KET_TM-BA-e-1420 9.2 Dispensing Place and tare full container. Remove portions of components and save them at the same time
KEN-TM / KET_TM-BA-e-1420 89 9.3 Checkweighing Checkweighing allows you to set an upper as well as a lower limit value (min / max threshold) a
KEN-TM / KET_TM-BA-e-1420 9 1.2 Anschlüsse / Schnittstellen 1.2.1 KET-TM 1. Ethernet 2. RS232 (COM1) 3. USB port 4. Digital I/O, RS232
90 KEN-TM / KET_TM-BA-e-1420 Entering limits: Press assignable function key (For assigning function to a key, see chap. 7.4.2). The numer
KEN-TM / KET_TM-BA-e-1420 91 Start checkweighing: Tare when using a weighing container. Put on goods to be weighed, tolerance control is sta
92 KEN-TM / KET_TM-BA-e-1420 9.4 Tare mode Tap < Tare mode >. Select tare mode; see table below. 8. Tab. 8: Subsequent tare weight
KEN-TM / KET_TM-BA-e-1420 93 9.5 Labelling The creation of labels is only possible in combination with label printers KERN PET A07 or PET A08.
94 KEN-TM / KET_TM-BA-e-1420 9.6 Statistics Statistics will be updated each time a new value is imported. This function is used to determi
KEN-TM / KET_TM-BA-e-1420 95 9.7 Differential weighing In differential weighing, samples are investigated with regard to changes in weight. The
96 KEN-TM / KET_TM-BA-e-1420 9.8 Peak value function This function shows the highest load value (peak value) in a series of weighing activit
KEN-TM / KET_TM-BA-e-1420 97 9.9 Info about saved weighing activities Activate function: Tap <Information about saved measurement>. Fun
98 KEN-TM / KET_TM-BA-e-1420 10 Parts counting Before the balance can count parts, it must know the average part weight (i.e. reference). Vari
KEN-TM / KET_TM-BA-e-1420 99 10.2 Application specific settings Tap icon in the bottom left corner of the window or call it up from the menu
Commentaires sur ces manuels